The primary purpose of the EF600-100 is to generate the cooling profile required for the cryopreservation of human embryos, in a sample plate designed to hold up to 16 x CBS high security straws. These specifications are successfully achieved
When fewer than 16 straws are processed they should be positioned in the centre of the sample plate; in this case the maximum straw to straw temperature difference is reduce to below ±0.5°C at all temperatures.
The temperate profile achieved within straws was not dependent on the loading; identical results were obtained when processing 1, 8 or 16 straws
In addition to the human embryo profiles, linear profiles for the cryopreservation of human embryonic stem cells and human sperm are successfully achieved
Control of specific non-linear profiles for the cryopreservation of embryos and oocytes is also possible.
|
Copyright©2008 Asymptote Ltd.